35.2642—Records of periodic spot-checks for teletherapy units.
        
        (a) 
         A licensee shall retain a record of each periodic spot-check for teletherapy units required by  § 35.642 for 3 years.
    
    
    
    
        
        (2) 
         The manufacturer's name, model number, and serial number of the teletherapy unit, source and instrument used to measure the output of the teletherapy unit;
    
    
    
    
        
        (5) 
         A determination of the coincidence of the radiation field and the field indicated by the light beam localizing device;
    
    
    
    
        
        (8) 
         Notations indicating the operability of each entrance door electrical interlock, each electrical or mechanical stop, each source exposure indicator light, and the viewing and intercom system and doors; and
    
    
        
        (9) 
         The name of the individual who performed the periodic spot-check and the signature of the authorized medical physicist who reviewed the record of the spot-check.
    
    
        
        (c) 
         A licensee shall retain a copy of the procedures required by  § 35.642(b) until the licensee no longer possesses the teletherapy unit.