1065.290—PM gravimetric balance.
        
        (b) Component requirements.
         We recommend that you use a balance that meets the specifications in Table 1 of  § 1065.205. Note that your balance-based system must meet the linearity verification in  § 1065.307. If the balance uses internal calibration weights for routine spanning and the weights do not meet the specifications in  § 1065.790, the weights must be verified independently with external calibration weights meeting the requirements of  § 1065.790. While you may also use an inertial balance to measure PM, as described in  § 1065.295, use a reference procedure based on a gravimetric balance for comparison with any proposed alternate measurement procedure under  § 1065.10.
    
    
        
        (c) Pan design.
         We recommend that you use a balance pan designed to minimize corner loading of the balance, as follows:
    
    
        
        (1) 
         Use a pan that centers the PM sample media (such as a filter) on the weighing pan. For example, use a pan in the shape of a cross that has upswept tips that center the PM sample media on the pan.
    
    
    
        
        (d) Balance configuration.
         Configure the balance for optimum settling time and stability at your location.