86.316-79—Carbon monoxide and carbon dioxide analyzer specifications.

(a) Carbon monoxide and carbon dioxide measurements are to be made with nondispersive infrared (NDIR) an analyzers.
(b) The use of linearizing circuits is permitted.
(c) The minimum water rejection ratio (maximum CO 2 interference) as measured by § 86.321 shall be:
(1) For CO analyzers, 1000:1.
(2) For CO 2 analyzers, 100:1.
(d) The minimum CO 2 rejection ratio (maximum CO 2 interference) as measured by § 86.322 for CO analyzers shall be 5000:1.
(e) Zero suppression. Various techniques of zero suppression may be used to increase readability, but only with prior approval by the Administrator.