89.316—Analyzer leakage and response time.
        
        (a) Vacuum side leak check.
        
        (1) 
         Any location within the analysis system where a vacuum leak could affect the test results must be checked.
    
    
        
        (2) 
         The maximum allowable leakage rate on the vacuum side is 0.5 percent of the in-use flow rate for the portion of the system being checked. The analyzer flows and bypass flows may be used to estimate the in-use flow rates.
    
    
        
        (3) 
         The sample probe and the connection between the sample probe and valve V2 (see Figure 1 in appendix B of this subpart) may be excluded from the leak check.