84.1145—Silica dust test; non-powered single-use dust respirators; minimum requirements.
        
        (b) 
         As described in  § 84.1144, airflow will be cycled through the respirator by a breathing machine at the rate of 24 respirations per minute with a minute volume of 40 liters; a breathing machine cam with a work rate of 622 kg.-m.2/minute shall be used.
    
    
    
    
        
        (e) 
         The total amount of unretained test suspension, after drying, in samples taken during testing, shall not exceed 1.8 milligrams for any single test.